Alicat P Series Pressure Gauge User Manual

Page 37

Advertising
background image

37

Changing From Streaming to Polling Mode:
When the meter is in the Streaming Mode (RS-485 units do not have a

streaming mode), the screen is updated approximately 10-60 times per second

(depending on the amount of data on each line) so that the user sees the data

essentially in real time. It is sometimes desirable, and necessary when using

more than one unit on a single RS-232 line, to be able to poll the unit.
In Polling Mode the unit measures the flow normally, but only sends a line of

data when it is “polled”. Each unit can be given its own unique identifier or

address. Unless otherwise specified each unit is shipped with a default address

of capital A. Other valid addresses are B thru Z.
Once you have established communication with the unit and have a stream of

information filling your screen:
1. Type *@=A followed by “Enter” (or using the RS-232 / RS-485

communication select menu, select A as identifier and exit the screen) to

stop the streaming mode of information. Note that the flow of information

will not stop while you are typing and you will not be able to read what you

have typed. Also, the unit does not accept a backspace or delete in the line

so it must be typed correctly. If in doubt, simply hit enter and start again. If

the unit does not get exactly what it is expecting, it will ignore it. If the line

has been typed correctly, the data will stop.

2. You may now poll the unit by typing A followed by “Enter”. This does an

instantaneous poll of unit A and returns the values once. You may type A

“Enter” as many times as you like. Alternately you could resume streaming

mode by typing *@=@ followed by “Enter”. Repeat step 1 to remove the

unit from the streaming mode.

3. To assign the unit a new address, type *@=New Address, e.g. *@=B. Care

should be taken not to assign an address to a unit if more than one unit

is on the RS-232 / RS-485 line as all of the addresses will be reassigned.

Instead, each should be individually attached to the RS-232 / RS-485 line,

given an address, and taken off. After each unit has been given a unique

address, they can all be put back on the same line and polled individually.

Advertising