Mounting, Plumbing – Alicat P Series Pressure Gauge User Manual

Page 7

Advertising
background image

7

MOUNTING

All P-Series Gauges and PC-Series Controllers have mounting holes for

convenient mounting to flat panels. These gauges are position insensitive and

can be mounted in any orientation. The sizes and dimensions for the mounting

holes are shown on pages 51 to 67.
P-Series Pressure Gauges may be connected into your system with the flow

going in either direction for ease of viewing the display. These units are shipped

with a plug for dead end applications. This plug should be removed for flow

through applications.
PC-Series Vacuum and Pressure Controllers are normally intended to control

the process pressure downstream of the controller. In order for this to occur the

controller should be mounted so the flow goes from left to right as you look at

the front of the unit. This puts the measuring portion of the device between the

valve and the leakage point where you are attempting to control the pressure
application. Back-pressure controllers reverse this configuration (see page 29).

PLUMBING

Your instrument is shipped with plastic plugs fitted in the port openings.

To lessen the chance of contaminating the flow stream do not remove

these plugs until you are ready to install the device.
Make sure that flow is in the direction indicated by the flow arrow.

Standard P-Series Gauges and PC-Series Controllers have female inlet and outlet

port connections. Welded VCR and other specialty fittings may have male ports.
The inlet and outlet port sizes (process connections) for different flow ranges

are shown on pages 51-67.
Instruments with M5 (10-32) ports have O-ring face seals and require no

sealant or tape. Do not use tape with welded or O-ring fittings.
For non M5 (10-32) ports use thread sealing Teflon® tape to prevent leakage

around the port threads.

Do not wrap the first two threads. This will minimize the possibility of getting

tape into the flow stream and flow body.

Do not use pipe dopes or sealants on the process connections as these

compounds can cause permanent damage to the controller should they get

into the flow stream.

When changing fittings, carefully clean any tape or debris from the port threads.

For additional notes on PCD (dual valve controller) plumbing see page 33.
For gas applications,
it is recommended that a 40 micron filter be installed

upstream of P and PCR-Series instruments and a 20 micron filter be installed

upstream of PC and PCD-Series instruments.
For liquid applications, see “Using Alicat Pressure Instruments with Fluids”, page 8.