A-100ne, 4 how to replace the pump tube, This page blank – Blue-White A-100N FLEXFLO Peristaltic Pump User Manual

Page 22: Page 23

Advertising
background image

A-100NE

Page 22

Remove the Old Pump Tube - The pump roller assembly spins in a
counter clockwise direction. The pump head inlet (suction) side is located
at the bottom of the pump and the outlet (discharge) is located at the top of
the pump head.

+ Release any pressure that may be in

the discharge tubing.

+ Disconnect the suction and discharge

tubes from the pump tube.

+ Remove the pump head cover.
+ With the pump running, pull the inlet

fitting out of the pumphead. Guide the
tube counter clockwise away from the
rollers. Pull the outlet fitting out of the
pump head.
Install the New Pump Tube - Be sure the pump head chamber is clean and
free of any debris.
Remove and inspect the roller assembly. Be sure the rollers spin freely. If
required, apply a small amount of grease to the pump head cover bearing.

+ With the pump running, insert the inlet (suction) side of the Pump Tube

fitting into the pump head.

+ Carefully guide the Pump Tube into the pump head. Stretch the tube

slightly and insert the outlet (discharge) fitting into the upper retaining slot
in the pump head.

+ Place the clear cover on the pump head and secure with three screws.

7.4 How to Replace the Pump Tube

The pump tube assembly will eventually break if not replaced. The tube has
been designed for a minimum service life of 500 hours. However, the life
of the tube is affected by many factors such as the type of chemical being
pumped, the amount of back pressure, the motor RPM, temperature and
others. The pump tube assembly must be inspected and replaced regularly.
After replacing the pump tube, press the Stand-by button and the Digit
button at the same time to reset the tube life warning timer.

Inlet Adapter

Outlet Adapter

Pump Head

RUN

STANDBY

RUN

FIELD

DIGIT

MODE

PROGRAM

STAND-BY

PRIME

MINIMUM

MAXIMUM

INPUT MODES

1 - MANUAL

2 - 4-20mA

3 - 0-10VDC

4 - 0-1000 Hz

PROGRAM

RESET SERVICE

PRIME

DISPLAY

DIGITAL TIMER PUMP

1000

MODE

5 - PULSE (BATCH)

A-100NE

Page 23

THIS PAGE BLANK

Advertising