Page 11

Sonic-Pro

The relative flow direction is shown on the Calibration Detail label affixed to the pipe fitting.

6.2.2.4
Fluid Flow
Direction

Each of the flow directions (A to B and B to A) have a Scaling Factor and either a positive or
negative Zero Factor associated with it. The correct factors are shown on the Calibration
Detail
label affixed to the ultrasonic flow sensor pipe fitting.

These are factory calibration factors that should not be changed without first
contacting the factory for technical assistance.

Transducer

Model

Mount Method

Cable Length(ft)

Flow Direction

Direction

B -> A

A -> B

Transducer

Model

Mount Method

Cable Length(ft)

Flow Direction

Settings

Direction

Zero(A->B)

Scaling(A->B)

Enter Value

Sign(A->B)

Zero Value(A->B)

Positive

B -> A

A -> B

Enter Value

Zero(B->A)

Scaling(B->A)

Enter Value

Sign(B->A)

Zero Value(B->A)

Enter Value

0.975

0.455

Negative

Positive

Negative

0.975

0.455

6.2.2.5
Fluid Flow
Direction

(CAL Settings)

6.2.3
Pipe
Menu

Pipe

Pipe Outside Dia.

Enter Value

Pipe Wall Thick.

Enter Value

Pipe Material

Custom

Pipe SOS

0.0000

Enter Value

0.0000

0.0000

HDPE

Each pipe fitting has a specific Pipe Outside Dia., Pipe Wall Thick., Pipe Material and Pipe
SOS
(Speed Of Sound) associated with it. The correct factors are shown on the Calibration
Detail
label affixed to the ultrasonic flow sensor pipe fitting.

These are factory calibration factors that should not be changed without first
contacting the factory for technical assistance.

CALIBRATION DETAIL LABEL

Page 12

Sonic-Pro

6.2.4
Fluid
Menu

Fluid

Fluid Type

Fluid SOS

Water

Custom

0000

Enter Value

The

sound transducers are located in the S4 pipe fitting. They are carefully positioned at the

factory for use with water at 68 degrees F which has a SOS (Speed Of Sound) of 1481
meters per second. (see page 28 for water SOS chart). The transducer placement cannot be
adjusted in the field. As the fluid’s sound speed moves away from the target Fluid SOS,
measurements become more difficult. If the speed that sound travels through the fluid is very
different than that of water, the meter will not function. The acceptable SOS range is from
1,400 to 1,600 m/sec.

The meter will display the SOS of the fluid being measured in the upper right hand corner of
the display.

Successful measurements are possible with fluids other than water provided the fluid’s SOS
is within the acceptable range and there are not too many particles or other sound blocking or
reflecting contaminants in the fluid which can inhibit the sound beam.

The target Fluid SOS value can be adjusted to maximize the meter’s ability to measure the
fluid. The chart below provides the sound speeds of a number of chemicals of specific
concentrations at specific temperatures. Diluted chemicals will have a different sound speed.
Select Custom from the Fluid Type menu, then input the SOS fluid in the Fluid SOS menu.

Fluid

Water

o

o

Temp. F ( C)

32 (0)

Sound Speed (meters/sec)

1403

Fluid Sound Speeds

Water

68 (20)

1481

Water

212 (100)

1543

1520

Sea Water

68 (20)

1900

Aluminum Sulfate 10% (Alum)

68 (20)

1578

Ethylene Glycol (50% in water)

77 (25)

1250

Gasoline

77 (25)

1581

Hydrochloric Acid 32%

1530

Hydrogen Peroxide 31%

1548

Milk

77 (25)

1250

Oil, Diesel

77 (25)

1487

Oil, Motor (SAE 30)

1431

Oil, Mineral

77 (25)

2800

Potassium Hydroxide 50%

68 (20)

68 (20)

68 (20)

68 (20)

1835

Sodium Hypochlorite 12%

68 (20)

1260

Sulfuric Acid 96%

68 (20)

Industries, Ltd.

Industries, Ltd.

ProSeries

by Blue-White Ind.

d

Industries, Ltd.

Industries, Ltd.

ProSeries

by Blue-White Ind.

d