Proseries, Sonic-pro, 3 contact closure relay output menu – Blue-White Sonic-Pro S4 Ultrasonic Flowmeters User Manual

Page 21

Advertising
background image

Page 19

Sonic-Pro

Page 20

Sonic-Pro

6.4.1
Analog
4-20mA
Output Menu

An analog current output signal can be programmed within the range of 4-20mA.

Specify the flow rate at a minimum current (Rate at Min Curr.) and the corresponding
minimum analog current (Min Analog O/P).

Specify the flow rate at a maximum current (Rate at Max Curr.) and the corresponding
maximum analog current (Max Analog O/P).

The flow rates must not be the same. The current for the high flow rate may be smaller than
the current for the low flow rate in which case, the current will decrease with increasing flow
rate.

Flow rates are mapped to currents using a straight line through the two points specified.

Analog 4-20mA

Rate at Min Curr.

0.0000

Enter Value

Min Analog O/P

4

Enter Value

Rate at Max Curr.

0.0000

Enter Value

Max Analog O/P

20

Enter Value

6.4.2
Pulse
0-1000 Hz
Output Menu

A digital pulse output signal can be programmed within the range of 0-1000 Hz (pulses per
second).

Specify the flow rate at a minimum frequency (Rate at Min Freq.) and the corresponding
minimum frequency (Min Frequency).

Specify the flow rate at a maximum frequency (Rate at Max Freq.) and the corresponding
maximum frequency (Max Frequency).

The flow rates must not be the same. The frequency for the high flow rate may be smaller
than the frequency for the low flow rate in which case, the frequency will decrease with
increasing flow rate.

Flow rates are mapped to frequencies using a straight line through the two points specified.

Pulse 0-1000

Rate at Min Freq.

0.0000

Enter Value

Min Frequency

0

Enter Value

Rate at Max Freq.

0.0000

Enter Value

Max Frequency

1000

Enter Value

6.4.3
Contact
Closure
Relay
Output Menu

The S4 has two Form C relays. Each can be independently configured to energize when one
of the following conditions are met:

!

A specific Flow Rate value is displayed, used for high, low or high/low range alarms.

!

An Accumulated Total value has been measured. Used for triggering external equipment

when a configured batch amount is reached.

!

A FAULT has occurred.

Contact Closure

Relay A

Monitor Flow Rate

High Trigger A

0

Enter Value

High Release A

0

Enter Value

Low Trigger A

0

Enter Value

Low Release A

0

Enter Value

Time(s) A

0

Enter Value

Monitor Totalizer

Batch Amt. A

0

Enter Value

Contact Timer(s) A

0

Enter Value

Faults A

Relay B

Monitor Flow Rate

High Trigger B

0

Enter Value

High Release B

0

Enter Value

Low Trigger B

0

Enter Value

Low Release B

0

Enter Value

Time(s) B

0

Enter Value

Monitor Totalizer

Batch Amt. B

0

Enter Value

Contact Timer(s) B

0

Enter Value

Faults B

None A

None B

Relay B

Monitor Flow Rate

High Trigger B

0

Enter Value

High Release B

0

Enter Value

Low Trigger B

0

Enter Value

Low Release B

0

Enter Value

Time(s) B

0

Enter Value

None B

6.4.3.1
CC Assigned to
Monitor Flow Rate

When in the Relay A or Relay B menu, selecting and then exiting the Monitor Flow Rate
menu will assigned that relay to monitor the rate of flow. The following functions are possible:

!

Set a High Trigger flow rate value which when measured, will energize the relay.

!

Set a High Release flow rate value which when measured, will de-energize the relay.

!

Set a Low Trigger flow rate value which when measured, will energize the relay.

!

Set a Low Release flow rate value which when measured, will de-energize the relay.

!

Set a Time (s) in seconds (0-999) to delay energizing the relay when it is triggered.

When either of the relays are energized, the display back-light will turn blue.

When both the trigger and release values are configured for the same value, the relay will
latch. The user must press the ESC button to clear a latched relay.