Page 17

Sonic-Pro

To conserve power, the S4 can be configured to “sleep” from 3 to 268 seconds per cycle. At
the end of each sleep cycle, the meter will “wake” for the number of seconds (from 2 to 10
seconds) as configured in the Flow Average feature, record data, and then return to sleep.
During the “wake” period, the display will remain blank.

The Flow Totalizer will estimate the amount of total fluid flow that would have been measured
while the meter was sleeping, using the following formula:

Total flow display value = T1 + (F1 + F2 / 2) x (N +1)
Where:
T1 = Last total before sleep
F1 = Last flow rate (units per second) before sleep
F2 = Flow rate average of the 1st second of wake-up (units per second)
N = Seconds of sleep (Sleep Timer Value)
note: the one second is added to the Sleep Time for accusation time at wake.

The sleep cycle can be interrupted by pressing the ENTER button.

Data logging cannot occur while sleeping. Logging will occur if the meter is “awake” and if it is
time to log data.

6.3.2.9
Sleep Timer

Sleep Time(s)

0

Enter Value

Page 18

Sonic-Pro

6.4
Configure the
SPU Output
Settings

In the Output Settings menu, the Analog 4-20mA, Pulse 0-1000 Hz, Contact Closures
and the Anybus Communications module functions can be configured.

Output Settings

Analog 4-20mA

Rate at Min Curr.

0.0000

Enter Value

Min Analog O/P

4

Enter Value

Rate at Max Curr.

0.0000

Enter Value

Max Analog O/P

20

Enter Value

Pulse 0-1000

Rate at Min Freq.

0.0000

Enter Value

Min Frequency

0

Enter Value

Rate at Max Freq.

0.0000

Enter Value

Max Frequency

1000

Enter Value

Contact Closure

Relay A

Monitor Flow Rate

High Trigger A

0

Enter Value

High Release A

0

Enter Value

Low Trigger A

0

Enter Value

Low Release A

0

Enter Value

Time(s) A

0

Enter Value

Monitor Totalizer

Batch Amt. A

0

Enter Value

Contact Timer(s) A

0

Enter Value

Faults A

Relay B

Monitor Flow Rate

High Trigger B

0

Enter Value

High Release B

0

Enter Value

Low Trigger B

0

Enter Value

Low Release B

0

Enter Value

Time(s) B

0

Enter Value

Monitor Totalizer

Batch Amt. B

0

Enter Value

Contact Timer(s) B

0

Enter Value

Faults B

None A

None B

Anybus

Type

Modbus RTU

Profibus DPV1

Modbus TCP

Ethernet
Profinet

None

Address

Modbus RTU

Profibus DPV1

Modbus TCP

Ethernet

Profinet

None

0

Enter Value

0

Enter Value

IP SET

192.168.234.123

OK

CANCEL

IP SET

192.168.234.123

OK

CANCEL

IP SET

192.168.234.123

OK

CANCEL