Proseries, Sonic-pro 6.3.2 display menu, Sonic-pro – Blue-White Sonic-Pro S4 Ultrasonic Flowmeters User Manual

Page 16: Page 15, Flow rate - volume units, Flow rate - time units, Flow rate - decimal point location, Totalizer - volume units, Totalizer - decimal point location, Flow rate averaging

Advertising
background image

Page 15

Sonic-Pro

6.3.2
Display
Menu

In the DISPLAY menu, the following features may be configured:

!

Flow Rate - Volume Units

!

Flow Rate - Time Units

!

Flow Rate - Decimal Point Location

!

Totalizer - Volume Units

!

Totalizer - Decimal Point Location

!

Flow Rate Averaging

!

Enable/Disable clearing the totalizer from the front panel

!

Low Flow Rate Cut-Off

!

High Flow Rate Cut-Off

!

Scaling Offset

!

Zero Offset

!

SOS Fault Window

!

Sleep Time

Display

Flow Rate

Volume Units

Ounces

U.S. Gallons

U.S. Barrels Liq.

U.S. Barrels Oil

Cubic Feet

Acre Feet

Imperial Gallons

Cubic Meters

Liters

Milliliters

FR Custom V/Gal

Time Units

Seconds

Hours

Days

Minutes

Fl. Rate Decimal

0

Enter Value

Totalizer

Volume Units

Ounces

U.S. Gallons

U.S. Barrels Liq.

U.S. Barrels Oil

Cubic Feet

Acre Feet

Imperial Gallons

Cubic Meters

Liters

Milliliters

TOT Custom V/Gal

Fl. Total Decimal

0

Enter Value

Flow Average(s)

0

Enter Value

Enable Tot. Clear?

Yes

No

Low Fl. Cut-Off

0

Enter Value

High Fl. Cut-Off

999

Enter Value

Scaling Off-set

1.0000

Enter Value

Zero Off-set

Sign

FR Offset

0.0000

Enter Value

Negative

Positive

SOS Fault Window

10

Enter Value

Sleep Time(s)

0

Enter Value

In the Flow Rate menu, the Volume Units and Time Units may be selected from a list.

A custom volume unit may be created by selecting FR Custom V/Gal and entering the
number of gallons in the custom volume.

The Fl. Rate Decimal decimal point location may be set to a maximum of 5 digits.

6.3.2.1
Flow Rate Menu

Flow Rate

Volume Units

Ounces

U.S. Gallons

U.S. Barrels Liq.

U.S. Barrels Oil

Cubic Feet

Acre Feet

Imperial Gallons

Cubic Meters

Liters

Milliliters

FR Custom V/Gal

Time Units

Seconds

Hours

Days

Minutes

Fl. Rate Decimal

0

Enter Value

0

Enter Value

0

Enter Value

0

Enter Value

Page 16

Sonic-Pro

In the Totalizer menu, the Volume Units may be selected from a list.

A custom volume unit may be created by selecting FR Custom V/Gal and entering the
number of gallons in the custom volume.

The Fl. Total Decimal decimal point location may be set to a maximum of 5 digits.

6.3.2.2
Totalizer Menu

Totalizer

Volume Units

Ounces

U.S. Gallons

U.S. Barrels Liq.

U.S. Barrels Oil

Cubic Feet

Acre Feet

Imperial Gallons

Cubic Meters

Liters

Milliliters

TOT Custom V/Gal

Fl. Total Decimal

0

Enter Value

0

Enter Value

Flow Average(s) sets the running average period in seconds. Any number of seconds from 2
to 10 seconds is possible.

6.3.2.3
Flow Average
Setting

Flow Average(s)

0

Enter Value

Enable Tot. Clear? enables or disables the ability of the user to clear the accumulated total
flow by pressing the ESC button.

6.3.2.4
Clear Total
Setting

Enable Tot. Clear?

Yes

No

Low Fl. Cut-Off sets the minimum displayed flow rate. Calculated flow rate values less than
this value will be displayed as Zero and flow totalizing will not occur.

6.3.2.5
Low Flow Rate
Cut-off Setting

Low Fl. Cut-Off

0

Enter Value

High Fl. Cut-Off sets the maximum displayed flow rate. Calculated flow rate values greater
than this value will be displayed as the High Fl. Cut-Off value, totalizing will still occur.

6.3.2.6
High Flow Rate
Cut-off Setting

High Fl. Cut-Off

999

Enter Value

Scaling Off-Set is a multiplier used to correct for flow inaccuracies. All flow rate values will
be multiplied by the Scaling Off-Set value and the result displayed and totaled.

6.3.2.7
Scaling Off-Set
Setting

Scaling Off-set

1.0000

Enter Value

Zero Off-Set is a factory calibration value used to offset the zero flow setpoint during the
initial calibration of the Ultrasonic Inline Pipe Fitting. The calibration value and sign should
match the ZERO: nnnn value and ZERO: +/-: xxx value shown on the Calibration Detail
label affixed to the pipe fitting.

These are factory calibration factors that should not be changed without first
contacting the factory for technical assistance.

6.3.2.8
Zero Off-Set
Setting

Zero Off-set

Sign

FR Offset

0.0000

Enter Value

Negative

Positive

The S4 will monitor the sound speed (SOS) of the measured fluid and report FAULT 12 if the
measured SOS exceeds the SOS Fault Window percentage value. The speed that sound
travels through the fluid will change if the density of the fluid changes (temperature change)
or if the fluid composition changes (chemical concentration, different chemicals, etc.). The
default value is 10%. The maximum allowable value is 50%.

6.3.2.8
SOS
(Speed Of Sound)
Fault Window

SOS Fault Window

10

Enter Value

CALIBRATION DETAIL LABEL

Industries, Ltd.

Industries, Ltd.

ProSeries

by Blue-White Ind.

d

Industries, Ltd.

Industries, Ltd.

ProSeries

by Blue-White Ind.

d