Proseries, Sonic-pro, Sonic-pro 6.4.4 anybus comm. menu – Blue-White Sonic-Pro S4 Ultrasonic Flowmeters User Manual

Page 22

Advertising
background image

Page 21

Sonic-Pro

Byte Bit

Type

No. of Bytes

1

0 Contact Closure 1

0: CC OFF, 1:CC ON

1 Contact Closure 2

0: CC OFF, 1:CC ON

2 Reserved
3 Reserved

1

4 Reserved
5 Reserved
6 Reserved
7 Reserved

2

0 Reserved
1 Reserved
2 Reserved
3 Reserved

1

4 Reserved
5 Reserved
6 Reserved
7 Reserved

3

Flow Rate

double

8

Current flow rate value.

11

Totalizer

double

8

Current totalizer value.

19

Fault Number Table

char

6

Fault number table upto six faults.

25

Speed of Sound

unsigned long

8

Current SOS value.

33

Goodness Of Meas.

unsigned char

1

Current GOM - 1 to 4.

34

Reserved

unsigned char

1

35

4 - 20mA Flow

unsigned char

1

4 to 20mA current output

36

Reserved

unsigned char

1

37

Flow Frequency

unsigned char

8

Frequency output

45

Firmware Version

unsigned char

8

8 ASCII characters. A11.22.33

53

Log Entry

unsigned char

86

Log Entry String
Eg: 2013/12/29 09:15:00 FR 0.0000 TL 0.00

Description

Page 22

Sonic-Pro

6.4.4
Anybus
Comm.
Menu

There are five different Anybus communications modules that can be installed in the S4
meter. Each optional module uses a different communications protocol.

6.4.3.1
CC Assigned to
Monitor Totalizer

When in the Relay A or Relay B menu, selecting and then exiting the Monitor Totalizer
menu will assigned that relay to monitor the accumulated to flow value. The following
functions are possible:

!

Set a Batch Amount total flow value which when reached, will energize the relay.

!

Set the Contact Timer(s) for the number of seconds (0-999) that the relay will remain

energized.

When either of the relays are energized, the display back-light will turn blue.

Monitor Totalizer

Batch Amt. B

0

Enter Value

Contact Timer(s) B

0

Enter Value

6.4.3.2
CC Assigned to
Monitor FAULTS

When in the Relay A or Relay B menu, selecting and then exiting the FAULTS menu will
assign that relay to energize when any of the following fault conditions exist.

Ÿ

FAULT 11 - Transducers are not connected or there is no fluid in the pipe.

Ÿ

FAULT 12 - The allowable SOS (Speed Of Sound) error set-point has been exceeded.

Ÿ

FAULT 21 - The configuration data is not valid.

The display back-light will turn red when there is a FAULT condition.

Faults B

Protocol

Anybus Communications Options

Profibus DPV

DB9F

Modbus RTU

Modbus TCP

Industrial Ethernet/IP

Profinet IO

Up to 12 Mbit/s

RJ45

10/100 Mbit/s full/half duplex

Network Connector

Baud Rate

DB9F

RJ45

RJ45

Up to 12 Mbit/s

10/100 Mbit/s full/half duplex

100 Mbit/s full duplex

Anybus

Type

Modbus RTU

Profibus DPV1

Modbus TCP

Ethernet
Profinet

None

Address

Modbus RTU

Profibus DPV1

Modbus TCP

Ethernet

Profinet

None

0

Enter Value

0

Enter Value

IP SET

192.168.234.123

OK

CANCEL

IP SET

192.168.234.123

OK

CANCEL

IP SET

192.168.234.123

OK

CANCEL

Select the module being used and enter the Port Address (for Profibus DPV or Modbus RTU
modules) or the IP ADDRESS ( for Modbus TCP, Industrial Ethernet/IP or Profinet IO
modules.

Use the following Anybus mapping structure to communicate with the module and monitor the
flowmeter. There are no send commands available.

Optional Anybus
Communications
Module

Communications
Cable Grommet

Industries, Ltd.

Industries, Ltd.

ProSeries

by Blue-White Ind.

d

Industries, Ltd.

Industries, Ltd.

ProSeries

by Blue-White Ind.

d