5.SETTING OF FUNCTIONS

34

5-5. Stability

Enter the parameters to detect stability.

If the difference between the current indicated value and the 2.5-msec-previously

indicated value is less than the set count and the duration of the condition is more than

the set time, the indicated value is regarded to be stable.

When stability is detected, a digital filter (fixed 32 times) to control instability in weight

value is automatically inserted. This stable-time digital filter differs from the digital filter

setting in the operation mode.

Setting range

- MD (time): 0.1 - 9.9 sec.

- MD (width): 01 - 99 markings

How to set

Set

Indicated Value

2.5msec

Time

Set Period

range

Setting call

Page 1

Operation Setting

CAUTION

When the time is 0.0 sec. and the width is 00 markings, stability is

not detected.

At this time, the stable-time digital filter is normally off.