5-3. valley hold, Valley hold – FUTEK IPM600 (D610) Digital Display User Manual

Page 69

Advertising
background image

7.HOLD FUNCTIONS

58

7-5-3. Valley Hold

The maximum value (valley) in the negative direction of the specified period is held.

The period is specified by the setting of “all period”, “external signal”, “time”, or “time

with trigger”.

(Example) All Period Valley Hold

t1 A delay time between the instant when the T/H signal is inputted and the instant

when the indicated value is held
1.0ms (max.)

t2 A delay time between the instant when the T/H signal is released and the instant

when the indicated value returns to tracking
1.0ms (max.)

t3

A minimum reset signal width required for releasing the hold

1.0ms (min.)

Indicated Value

t2

t3

Sensor Input Value

Detection

t

T/H

HI-LO Limit
Judging Output

H/E

Detection Hold Period

t1

t2

t1

t3

Note: During the undetermined period, the judging output varies with

fluctuations in the input waveform. However, the H/E output

remains on during the undetermined period. Read the judging

result when the indicated value becomes stable (immediately

before the T/H signal rises).

OFF

ON

OFF

ON

OFF

ON

OFF

ON

HOLD

Status Display

PEAK

Status Display

ON

Blink

Blink

OFF

ON

OFF

OFF

OFF

OFF

OFF

HOLD
Button

Undetermind
Period Note

Undetermind
Period Note

Hold Period