ROTRONIC CF8-W-Disp-CO User Manual

Page 8

Advertising
background image

Dokument
UMA 147

Rev
1

Page
8 (16)

6d1

The present
MIN value in
% of FS

no

Increase the MIN value by stepping with the MENU(+) button.
Decrease by stepping with the ESC(-) button. The output is set to
the MIN value and can be checked with a multimeter.

6e1

Sure

no

The adjustment is saved by pressing Entr(+-). Press ESC to return
without saving.

6c2

SetH

no

Leads to adjustment of the MAX value

6d2

The present
MAX value in
% of FS

no

Increase MAX value by stepping with the MENU(+) button. Decrease
by stepping with the ESC(-) button. The output is set to the MAX
value and can be checked with a multimeter.

6e2

Sure

no

The adjustment is saved by pressing Entr(+-). Press ESC to return
without saving.

7

CALb

yes

Background calibration of the CO sensor and the CO

2

-sensor

with fresh air. An easy way to correct the zero point error. The
sensor needs fresh air (380-420 ppm CO

2

). The calibration must be

confirmed by pressing Entr(+-). The CO

2

sensor is calibrated to 400

ppm CO

2

and the CO sensor to 0 ppm CO.

7a

Sure

yes

Confirm that a background calibration shall be done.

8

CAL

yes

Zero point calibration of the CO sensor and the CO

2

sensor. The

sensor needs zero gas. See the zero point calibration instruction.
The calibration must be confirmed by pressing Entr(+-).

8a

Sure

yes

Confirm that a zero point calibration shall be done.

Table IV. Maintenance functions available on CF8-W-Disp-CO

to set output limits. Time limit

refers to an internal time-out that returns the LCD and maintenance function back to normal.
ENTER is a simultaneous pressing on MENU and ESC.


PLEASE NOTE!
If a power failure happens when the sensor has Out(1)...(4) locked
to min/max limits, then the sensor will have this output locked when the power
returns. It is necessary to enter this menu item and release the output manually!