Setting of the max value of the analogue outputs – ROTRONIC CF8-W-Disp-CO User Manual

Page 9

Advertising
background image

Dokument
UMA 147

Rev
1

Page
9 (16)

EXAMPLE I:


Setting of the MAX value of the analogue outputs

The access code is time limited. If the time limit is exceeded the sensor returns to
DEFAULT.

ENTR =MENU(+) and ESC(-) are pushed simultaneously.

1.

At the start of the setting the sensor is in DEFAULT.

2.

Push MENU(+) once to reach the error code. The display shows E + the error
code number. No error is shown as E000.

3.

Push MENU(+) once. The display shows the CO set point e.g. 35.


4.

Push ENTR(+-) once. The display shows EC and two digits. Enter the access
code to the service menu. If the default code is used do like this: Push
MENU(+) until the digits stop. The display shows EC55. If the correct code is
not entered before the time limit exceeds the sensor returns to DEFAULT.


5.

Push ENTR(+-) once. The display shows SPCO to set the CO set point.


6.

Push MENU(+) once to reach the carbon dioxide set point. The display shows
SP C.


7.

Push MENU(+) once to reach the setting of analogue outputs. The display
shows AOUt.


8.

Push ENTR(+-) to reach the output to be set. The display shows An and two
digits e.g. An 01. Step to the requested output by pushing MENU(+).


9.

Push ENTR(+-) to reach the setting of the MIN value. The display shows Set
L.


10.

Push MENU(+) to reach the setting of the MAX value. The display shows Set
H.


11.

Push ENTR(+-) to set the MAX value of the output. The display shows the
numerical value in % e.g. the standard setting100.0 % or previously set value.
Push MENU(+) to increase the MAX value. Push ESC(-) to decrease the MAX
value. Push ENTR(+-). The display shows Sure. Push ENTR(+-) to save the
setting and return to Set H. Push MENU(+) or ESC(-) to return to Set H
without saving the new setting. Push ESC(-) to return to the output to be set,
item 12. Push ESC(-) once again to return to the setting of analogue outputs
item 11.