6.3 Composition of an RTU request message (transmitted from a high-level

computer to this product)

 For codes a) through i), see "6.5 Description of RTU codes."

6.3.1 Composition of a read request message

a) Slave address

01H

b) Function code

03H

c) Register address

High level

00H

First register address

Low level

00H

d) Number of registers

High level

00H

Fixed at 2

Low level

02H

e) CRC-16

Low level

0BH

High level

C4H


6.3.2 Composition of a write request message

a) Slave address

01H

b) Function code

10H

c) Register address

High level

00H

First register address

Low level

00H

d) Number of registers

High level

00H

Fixed at 2

Low level

02H

f)

Number of byte items

04H

Number of registers

2

g)

Data for the first register
(a low-level word)

High level

00H

Data composition is ①, ②, ③,
and ④H. (① represents 1
byte.)

Low level

00H

Data for the first register
(a high-level word)

High level

00H

Low level

00H

e) CRC-16

Low level

3FH

High level

FEH


6.3.3 Composition of a store request message

a) Slave address

01H

b) Function code

10H

c) Register address

High level

02H

First register address

Low level

0EH

d) Number of registers

High level

00H

Fixed at 2

Low level

02H

f)

Number of byte items

04H

Number of registers

2

g)

Data for the first register
(a low-level word)

High level

00H

The data about the storage of settings
is arbitrary.

Low level

00H

Data for the first register
(a high-level word)

High level

00H

Low level

00H

e) CRC-16

Low level

E2H

High level

EBH

Advertising