General Information 13

Table 1-2. Supplemental Characteristics (continued)

Agilent Model

Parameter

66101A

66102A

66103A

66104A

66105A

66106A

Temperature Coefficients (change in
output per

°

C after a 30-minute

warmup)

Output Voltage:

30 ppm +

0.1 mV

0.2 mV

0.3 mV

0.6 mV

1.1 mV

1.8 mV

Output Current:

30 ppm +

0.2 mA

0.1 mA

0.05 mA

0.03 mA

0.01 mA

0.006 mA

Voltage Readback:

20 ppm +

0.1 mV

0.2 mV

0.2 mV

0.7 mV

0.7 mV

1.1 mV

±

Current Readback:

20 ppm +

0.2 mA

0.1 mA

0.05 mA

0.02 mA

0.02 mA

0.02 mA

OV Protection:

1 mV

2 mV

4 mV

8 mV

16 mV

30 mV

Maximum Reverse Current
(maximum current, without damage,
that the power module will withstand
while turned on and with the dc output
reverse biased by an external dc
source):

16 A

7.5 A

4.5 A

2.5 A

1.25 A

0.75 A

Typical Output Common Mode Noise
Current*
rms:

0.5 mA

peak-peak:

5 mA

* Referenced to mainframe chassis ground.

Remote Sensing Capability
Voltage drop per lead:

Up to 1/2 of rated output voltage

Load voltage:

Subtract voltage drop in load leads from specified output voltage rating

Load Regulation:

For each 1-volt change in the negative output lead due to load current change,

add 2 mV to the voltage load regulation specification (see Table 1-1).

Command Processing Time (average
time for output to begin to change after
receipt of digital data when the power
module is connected directly to the
GPIB):

20 ms

Output Voltage Programming
Response Time*
Time for output to change from 10% to
90% of its total excursion:

< 20 ms

Time for output to change from 90% to
10% of its total excursion:

< 20 ms

< 50 ms

Time for output voltage to settle to
within 0.1% of the final value:

120 ms

* Excluding command processing time