28 Turn-On Checkout

Power Module Turn-On State

When you turn on the mainframe, the power module goes through a self-test. All front panel readout digits and
annunciators turn on briefly. The panel then remains in the state shown in Table 3-1A.

Note

The figure assumes the power module still retains its factory default turn-on state. You can change this
state if desired (see "Chapter 4 - Basic Power Module Commands").

Checking Basic Module Functions

The power module is designed for remote programming over the GPIB. It also can be locally programmed from the
optional Agilent 66001A MPS Keyboard. Table 3-2 gives the SCPI commands for the remotely programmed tests. You
can send these same commands from the keyboard (see "Appendix B").

Table 3-2. Testing Basic Functions with SCPI Commands

Command

Action

Display Response

VOLTS

AMPS

CV

CC

Addr*

Dis

Prot

Output Voltage (Disconnect Load)

VOLT 5.1

Set output voltage to 5.1 V

0.00

0.00

Off

Off

On

On

Off

OUTP ON

Enable the output

5. 10

0.00

On

Off

On

Off

Off

Overvoltage Protection

VOLT:PROT 4.9

Activate OVP circuit (by
setting protection voltage

0.00

0.00

Off

Off

On

Off

On

below output voltage)

VOLT:PROT MAX

Clear OVP circuit

5. 10

0.00

Off

Off

On

Off

Off

OUTP:PROT:CLE

Clear OVP circuit

5. 10

0.00

On

Off

On

Off

Off

Save and Recall Functions

*SAV 5

Save present state to
location 5

5.10

0.00

On

Off

On

Off

Off

VOLT 3.55

Set output voltage to 3.55 V

3.55

0.00

On

Off

On

Off

Off

OUTP OFF

Turn off the output

0.00

0.00

Off

Off

On

On

Off

*SAV 6

Save present state to
location 6

0.00

0.00

Off

Off

On

On

Off

*RCL 5

Restore state 5

5.10

0.00

On

Off

On

Off

Off

*RCL 6

Restore state 6

0.00

0.00

Off

Off

On

On

Off

Output Current (With output disabled, short the
output terminals with a wire size sufficient to carry
the maximum current of the module.)

CURR 3.1

Set output current to 3.1
amps

0.00

0.00

Off

Off

On

On

Off

OUTP ON

Enable the output

0.00

3.10

Off

On

On

Off

Off

Overcurrent Protection

CURR:PROT:STAT
ON

Activate OCP circuit (by
enabling OCP when the
output is shorted)

0.00

0.00

Off

Off

On

Off

On

CURR:PROT:STAT
OFF

Clear OCP circuit

0.00

0.00

Off

On

On

On

Off

OUTP:PROT:CLE

Clear OCP circuit

0.00

3.10

Off

On

On

Off

Off

*Addr will not light if the module is controlled from the keyboard.