Basic Power Module Commands 33

Controlling List-Mode Output

The following commands control the power module list-mode voltage and current output. Lists require a trigger to begin
execution. For an explanation of lists and triggers, see "Chapter 5 - Synchronizing Power Module Output Changes": in the
Programming Guide.

Table 4-4. Power Module List-Mode Output Commands

Command

Function

CURR:MODE LIST

Sets the current mode to list (as opposed to fixed).

LIST:CURR <value>,<value>

Programs the output current values (or points) in the list. When you execute the list,
the output will sequentially step through the list values.

LIST:CURR:POIN?

Returns the number of points programmed in the current list.

VOLT:MODE LIST

Sets the voltage mode to list (as opposed to fixed).

LIST:VOLT <value>,<value>

Programs the output voltage values (or points) in the list. When you execute the list,
the output will sequentially step through the list values.

LIST:VOLT:POIN?

Returns the number of points programmed in the voltage list.

LIST:DWEL <value>,<value>

Programs the dwell value, in seconds, for each value in a voltage or current list.

There must be one dwell value for each value in the voltage or current list.

LIST:DWEL:POIN?

Returns the number of points programmed in the dwell list.

LIST: STEP ONCE

Commands the list to execute only one value (point) when a trigger is received. This
specifies a trigger-paced list.

LIST: STEP AUTO

Commands the list to execute all values (points) when a trigger is received. This
specifies a dwell-paced list.


Controlling Triggers

The following commands control the power module trigger subsystem. For a detailed description of this subsystem, see
"Chapter 5 - Synchronizing Power Module Output Changes": in the Programming Guide.

Table 4-5. Power Module Trigger Commands

Command

Function

INIT

Initiates the trigger subsystem to detect one trigger. No triggers will be processed
unless the subsystem is initiated.

INIT:CONT

Keeps the trigger subsystem continuously initiated so it will process each incoming
trigger.

TRIG:SOUR
BUS|EXT|HOLD|LINK| TTLT

Specifies which trigger source is to be accepted by the trigger subsystem (see the
Programming Guide). HOLD prevents response to all triggers.

TRIG:DEL <value>

Programs the delay time (in seconds) between the detection of a trigger and its
execution.

TRIG *TRG

Either command sends a trigger signal to the power module over the GPIB bus.

TRIG:LINK <parameter>

When TRIG:SOUR LINK is programmed, this command specifies the link parameter
for the trigger (see Programming Guide).

OUTP: TILT ON

Enables the power module Trigger Out signal, which is a 20-

µ

s negative-true TTL

pulse available at the mainframe TRIGGER OUT jack. To use this output, you must
program a TTLT trigger source (OUTP:TTLT:SOUR).

OUTP: TTLT OFF

Disables the power module Trigger Out signal.

OUTP:TTLT:SOUR
BUS|EXT|HOLD|LINK

Specifies which trigger source is to be used for the Trigger Out signal. HOLD
prevents a response to any source.

OUTP:TTLT:LINK

When OUTP: SOUR LINK is programmed, this command specifies the link
parameter for the trigger (see Programming Guide).