20 Installation

Installing The Module In The Mainframe

Figure 2-3. Power Module Installation

Note

A fully loaded mainframe can weigh over 36 kg (80 lbs). Install the mainframe into the rack (see the
"Mainframe Users Guide") before installing the modules into the mainframe.

1.

Turn off the mainframe power switch.

2.

Make certain the pull tab extends from the module (see Figure 2-3).

3.

Slide the module into the mainframe slot. You may use any available slot.

4.

Push the module into the backplane connector until you feel the module connector snap into the backplane connector.

5.

Push the lock tab into the front of the module. If the module is not properly plugged into the backplane connector, the
lock tab will not retract into the module.

6.

If desired, mark or label the power module with its slot address.

Important

Marking or labeling the power module will prevent replacing it in the wrong slot if it is subsequently
removed from the mainframe. Also, with two mainframes linked together, a slot can be either of two
numbers, depending on whether it is in the main or the auxiliary mainframe (see "Connecting the
Controller" in this chapter).

Connecting The Load

Output Connector

One output connector (see Table 1-3) is shipped with each power module. This can be either a standard or relay type. The
Installation Guide supplied with the connector shows how to open the connector, connect the wires, and secure the
connector to the mainframe.