Agilent Technologies 66001A User Manual

Page 14

Advertising
background image

14 General Information

Table 1-2. Supplemental Characteristics (continued)

Agilent

Model

Parameter

66101A 66102A 66103A 66104A 66105A 66106A

Trigger Response Time (average time
for the output voltage or current to
begin to change upon receipt of an
external trigger)

With GPIB activity:

5 ms

Without GPIB activity:

3 ms

Trigger Maximum Frequency (for
repetitive external triggering, the
maximum frequency for which the
Trigger Response Time is valid)

With GPIB activity:

75 Hz

Without GPIB activity:

85 Hz

Dwell Time

Programmable range:

10 ms to 65 s

Resolution:

2 ms

Accuracy

With GPIB activity:

± 5 ms

Without GPIB activity:

± 2.5 ms

Downprogramming:

Active downprogrammer sinks approximately 10% of rated output current

Monotonicity:

Output is monotonic over entire rated voltage, current, and temperature range

GPIB Interface Capabilities

Interface Signals (see IEEE 488.1):

AH1,C0,DC1, DT1,EI,LE4,PP0, RL1,SH1,SR1,TE6

Programming Language:

SCPI (Standard Commands for Programmable Instruments)

Serial Link Capabilities (maximum
power modules sharing one GPIB
primary address)

16

Savable States

10

Nonvolatile memory locations:

5 (0 through 4)

Volatile memory locations:

5 (5 through 9)

Nonvolatile memory write cycles:

40,000 (typical)

Recommended Calibration Interval:

1 year

Safety Compliance

Complies with:

CSA 22.2 No.231 & IEC 348

Designed to comply with:

UL 1244 & VDE 0411

RFI Suppression (designed to comply
with):

FTZ 1046/84, Level B

Dimensions

Width:

48.3 mm (1.9 in)

Height:

141.2 mm (5.6 in)

Depth

Pull tab inserted:

571.4 mm (22.5 in)

Pull tab extended:

606.4 mm (23.8 in)

Weight

Net:

2.7 kg (6 lb)

Shipping:

4.1 kg (9 lb)

Product Storage Temperature

-40° C to +70° C

Advertising