Veriteq Spectrum 4.0 User’s Guide

31

5. For Timer Value, set the number of minutes or seconds you want the trans-

ducer to begin warming up before taking a sample. The default is 10 sec-
onds.

6. Click OK.

Configuring Data Logger Channels to Work with Thermocouples

This section applies only to those using Veriteq SP-1700 Series Loggers. Ver-
iteq SP-1700 loggers work with various types of thermocouples (E, J, K, R, S,
and T). You need to configure your channel to specify the type of thermocouple
you are using.

To configure SP-1700

loggers to work with thermocouples:

1. In Spectrum, choose Logger>Setup.

2. From the Setup window, click Setup next to the channel to configure.

3. From the Setup Thermocouple Channel screen, select Enable Channel.

4. From the Display Method list, choose your thermocouple type.

5. Enter a description.

6. Advanced: If your installation includes very long thermocouple cables,

Veriteq may recommend you specify a temperature offset. Enter it here. All
values will be adjusted by this offset.