Vaisala vLog 4.5 User Manual

Page 35

Advertising
background image

Configuring Loggers

Vaisala Veriteq vLog 4.5

29

Logger Status

Stopped at specified time: The logger is not sampling and

has stopped because the user-set stop time has been
reached.

Stopped when full: The logger is not sampling and has

stopped because it is configured to stop when the memory is
full.

Start at specified time: The logger is not sampling, but is set

to begin sampling as soon as the Start time is reached.

No enabled channels: The logger is not sampling because no
channels are enabled.

The Setup button to the right of the Logger Status enables you to
clear the logger’s memory and change the Sample Interval,
Start Mode, and Stop Mode settings.

Start Time

Indicates either the time that the currently-connected logger
began taking samples, or the time (in the future) that it is set to
begin taking readings
Future Start times are indicated by a (

) symbol.

For information on changing the Start Time, see “Setting the
Start Mode” on page 34.

Sample
Interval

Indicates the current sample interval setting (how frequently the
logger is programmed to take readings). For information on set-
ting sample intervals, see “Setting the Sample Interval” on
page 37.
• Note that the most frequent sampling rate possible with

VL-1000-VLT, VL-1016, VL-1200 and SP-1200, SP-1016, VL-
1416 and SP-1416 loggers is once per minute. For other log-
gers, the most frequent sampling possible is once every 10
seconds; however, sampling of this frequency will have a neg-
ative effect on battery life.

Stop Mode

Indicates how the currently-connected logger is set to stop while
recording samples in its memory. There are three choices:

Wrap

when full

,

Stop when full

, and

Stop at specified time

. For more

information on setting Stop modes, see “Setting the Stop Mode”
on page 35.

Table 3: Logger Setup window information

Advertising