Setting Sample Smoothing

Vaisala Veriteq vLog 4.5

79

2

Under Temperature scale, select

Fahrenheit

or

Celsius

.

This setting affects the data in all graphs, displays, and
printed reports.

Setting Sample Smoothing

In vLog, sample smoothing controls how samples in a logger
file are interpreted when they are inserted into a graph,
allowing you to dampen the effects of ‘noise’ in Logger files.
Sample smoothing is the minimum change in input signal
required to cause a change in displayed channel values, and
is defined in raw analog-to-digital converter counts.
vLog allows you to set two different sample smoothing
values: one for thermocouple channels, the other for other
channel types.

Note:

Sample smoothing only affects how samples are
interpreted and presented in graphs and reports;
the data in the loggers and the logger files is not