Configuring Loggers

Vaisala Veriteq vLog 4.5

39

Enabling and Disabling Logger Channels

The

Logger Channels

option permits you to enable or disable

the channels on the currently-connected logger.

Note:

During this process the logger’s memory is
cleared. If the information on the logger is still of
value, transfer the data to a PC before enabling
or disabling channels.

To enable or disable logger channels:
1

Select

Logger | Channels

.

2

In the window that appears, select or deselect the
checkbox to enable or disable a channel.

Click

OK

. This initiates the process of clearing the log-

ger’s memory. To conserve logger memory it is a good
idea to disable unused channels.

Note:

If you are using a temperature and humidity log-
ger, remember that the humidity sensor requires
temperature values; therefore, you cannot dis-
able the temperature sensor channel if you are
using the humidity channel.

Configuring Channels to Work with Transducers

This section applies only to those using Vaisala Veriteq VL-
4000 and SP-4000 Series Data Loggers. Used with
transducers, you can configure VL-4000 and SP-4000 Series
loggers to display measurements in units other than
milliAmps or volts.

To configure VL-4000 and SP-4000 Series Data Loggers to work
with transducers:
1

In vLog, choose

Logger | Setup

( ).

2

From the Setup window, click

Setup

next to the channel

to configure. Different loggers offer different channel
types, each with different setup windows and slightly