External Trigger Out

Trigger Out selects one of the eight lines from trigger bus
(TRIG [0..7]) as an external trigger source. The selection of
the trigger out line is done by the USB device in the chassis
as illustrated in

Figure 2- 2

by means of a multiplexer.

Table 2- 2

defines the available trigger out signals provided

by U2300A, U2500A, and U2600A Series DAQ, whereas for
U2700A Series modular products, user is allowed to choose
any trigger line from the trigger bus (TRIG [0..7]) as an
external trigger source.

The SCPI command below is used to select one of the lines
or bits of the trigger bus (TRIG [0..7]) as an external trigger
source:

TRIGger:OUT {0|1|2|3|4|5|6|7}

Table 2-2 Trigger out bits for U2300A, U2500A, and U2600A Series DAQ

devices

Trigger Out

Function

Bit-0

Time base

Bit-1

Reserved

Bit-2

Reserved

Bit-3

A/D trigger

Bit-4

Reserved

Bit-5

Reserved

Bit-6

Reserved

Bit-7

D/A trigger