Table 2-4 Example of configurations for multiple Master–multiple Slaves

M — Master, S — Slave, T0~T7 — Trigger bus (TRIG [0..7]), * — Star Trigger

[1]

Same trigger line is not allowed for multiple Master configuration.

[2]

Slave device not allowed to occupy two trigger lines.

[3]

Not allowed to have both Master and Slave configuration for a device.

[4]

Not allowed to have Star Trigger and Slave mode for a device.

Slot 1

Slot 2

Slot 3

Slot 4

Slot 5 & Slot 6

U2701A

U2702A

U2761A

U2751A

U2722A

Supported configurations

Configuration 1

M = T0

S = T0

S = T0

None

S = T0

Configuration 2

S = T1

M = T1

None

None

S = T1

Configuration 3

M = T0

M = T1

S = T0

None

S = T1

Configuration 4

*(Out)

M = T1

S = T1

* (In)

None

S = T1

Configuration 5

*(Out)

* (In)

* (In)

None

* (In)

Not supported configurations

Configuration 1

[1]

M = T0

M = T0

S = T0

None

S = T0

Configuration 2

[2]

M = T3

S = T3
S = T4

M = T4

None

S = T4

Configuration 3

[3]

M = T0

S = T1

S = T0

S = T0

None

S = T1

Configuration 4

[4]

*(Out)

M = T1

* (In)

S = T1

None

None

None