U2781A User’s Guide

2

Features and Functions

Simultaneous Synchronization (SSI)

Simultaneous Synchronization (SSI) provides synchronization
between the modular products within the chassis.

Figure 2- 4

illustrates an example of SSI. The SSI feature should be
configured using the bundled Agilent Measurement Manager
(AMM).

SSI allows users to set the modules as MASTER or SLAVE.
The MASTER module sends the SSI signal to the slave
modules via the backplane trigger bus (TRIG [0..7]). SLAVE
modules will then receive the signal and begin
synchronization with MASTER module.

There are two SSI configuration modes available — single
Master–multiple Slaves and multiple Masters–multiple Slaves.

N O T E

Only ONE master can be assigned for U2300A, U2500A, and U2600A Series.

For more information, refer to the AMM Help File, Chassis Trigger page.