adjustment factor =

input range

output

=

50 mV
55 mV

=

.909

The excitation adjusted to output 50 mV is:

adjusted excitation

= (original excitation)*(calculated adjustment factor)

= (5000 mV)*(.909)

= 4545 mV

Use 4545 mV to excite the sensor instead of 5000 mV within either the BrFull
or BrFull6W instruction.

6. CRBasic Program Examples

6.1 Wiring for the Program Examples

6.1.1 BrFull Instruction (4-Wire Full Bridge Measurement)

Datalogger

CS420-L/CS425-L

Excitation

E1

Red (Supply +)

Input

H1

Yellow (Output +)

Input

L1

Blue (Output -)

Analog Ground

AG

White (Supply -)

G

Clear

(Shield)

The orange and black wires are not used and the ends of these wires should be
taped to prevent shorting.

6.1.2 BrFull6W Instruction (6-Wire Full Bridge Measurement)

Datalogger

CS420-L/CS425-L

Excitation

E1

Red (Supply +)

Input

H1

Orange (Monitor +)

Input

L1

Black (Monitor -)

Input

H2

Yellow (Output +)

Input

L2

Blue (Output -)

An. Ground

AG

White (Supply -)

G

Clear

(Shield)