Edlog program examples, 1 cr10(x), cr510, and cr500 examples, 1 instruction 6 (4-wire full bridge measurement) – Campbell Scientific CS420 and CS425 PDCR 1830-8388 Pressure Transducer User Manual

Page 15: 1 wiring, 2 edlog program

Advertising
background image

CS420-L and CS425-L Pressure Transducers

7. Edlog Program Examples

7.1 CR10(X), CR510, and CR500 Examples

7.1.1 Instruction 6 (4-Wire Full Bridge Measurement)

7.1.1.1 Wiring

Datalogger

CS420-L/CS425-L

Excitation

E1

Red (Supply +)

Input

H1

Yellow (Output +)

Input

L1

Blue (Output -)

An. Ground

AG

White (Supply -)

G

Clear

(Shield)

The orange and black wires are not used and the ends of these wires should be
taped to prevent shorting.

7.1.1.2 Edlog Program

; Measure one 4-wire DRUCK Sensor.

* Table 1 Program
01: 5

Execution

Interval

(seconds)


1: Full Bridge (P6)
1: 1

Reps

2: 3

25 mV slow range

3: 1

DIFF

channel

4: 1

Excite all reps w/Exchan 1

5: 2500

mV

excitation

6: 1

Loc

[DEPTH_FT]

7: 2.3529 Mult
8: -0.49

Offset


; Every 60 minutes output the array ID, time,
; and average water depth in feet

2: If Time Is (P92)
1: 0

Minutes (seconds --) into

2: 60

Interval (same units as above)

3: 10

Set output flag high (Flag 0)


3: Real Time (P77)
1: 110

Day, Hour/Minute (midnight = 0000)


4: Average (P71)
1: 1

Reps

2: 1

Loc

[DEPTH_FT]

11