4 edlog example – Campbell Scientific CS420 and CS425 PDCR 1830-8388 Pressure Transducer User Manual

Page 23

Advertising
background image

CS420-L and CS425-L Pressure Transducers


LoopCount = LoopCount + 1

NextSubScan

PortSet (1 ,0)' Deactivate AM16/32


CallTable

WellData

NextScan
EndProg

8.1.4 Edlog Example

The following program is for the CR10(X); a CR23X is programmed similarly.

Each sensor has an independent multiplier and offset that must
be applied. To determine multiplier (M) and offset, see the
Multiplier and Offset sections.

NOTE

* Table 1 Program
01: 60

Execution

Interval

(seconds)


; Set Reset on AM16/32A Multiplexer:

1: Do (P86)
1: 41

Set Port 1 High


2: Beginning of Loop (P87)
1: 0000

Delay

2: 4

Loop

Count


3: Do (P86)
1: 72Pulse Port 2

4: Excitation with Delay (P22)
1: 1

Ex

Channel

2: 0

Delay w/Ex (units = 0.01 sec)

3: 1

Delay after Ex (units = 0.01 sec)

4: 0

mV

Excitation


; Measure Druck Pressure Transducer

5: Full Bridge w/mV Excit (P9)
1: 1

Rep

2: 5

2500 mV Slow Ex Range

3: 3

25 mV Slow Br Range

4: 1

DIFF

Chan

5: 1

Excite all reps w/Exchan 1

6: 2500

mV

Excitation

7: 1--

Loc [DEPTH_FT] (--Index Input Loc.1-4)

8: 1

Mult

9: 0

Offset