CS420-L and CS425-L Pressure Transducers

7.2 CR23X and 21X Examples

7.2.1 Instruction 6 (4-Wire Full Bridge Measurement)

7.2.1.1 Wiring

Datalogger

CS420-L/CS425-L

Excitation

E1

Red (Supply +)

Input

H1

Yellow (Output +)

Input

L1

Blue (Output -)

G

White (Supply -)

G

Clear

(Shield)

The orange and black wires are not used and the ends of these wires should be
taped to prevent shorting.

7.2.1.2 Edlog Program

; Measure one 4-wire DRUCK Sensor.

* Table 1 Program
01: 5

Execution

Interval

(seconds)


1: Full Bridge (P6)
1: 1

Reps

2: 22*

50 mV, 60 Hz Rejection, Slow Range

3: 1

DIFF

Channel

4: 1

Excite all reps w/Exchan 1

5: 5000

mV

Excitation

6: 1

Loc

[DEPTH_FT]

7: 2.3529 Mult
8: -0.49

Offset

*for 21X use Range Code 3

; Every 60 minutes output the array ID, time,
; and average water depth in feet.

2: If time is (P92)
1: 0000

Minutes

into

a

2: 60

Minute

Interval

3: 10

Set output flag high


3: Real Time (P77)
1: 110

Day, Hour/Minute (midnight = 0000)


4: Average (P71)
1: 1

Reps

2: 1

Loc

[DEPTH_FT]

13