Section 1. Product Overview

1-12

1.3.4 Controller Specifications

TABLE 1-7. Controller Specifications

Feature Function

Capability

START DELAY

Disabled

No start delay.

Time/Day

Adjustable, up to 1 week in advance.

Pulse Count

Adjustable, up to 9,999,999.

4-20mA

Adjustable, up to 9,999,999 (4-20mA = 0-100 Pulses/min).

External Contact

Momentary, 25 millisecond dry contact closure.

Level Control

Adjustable up to 99 second contact duration.

SAMPLE INITIATION

Disabled

No sample initiation.

Interval Time

Adjustable up to 999 hours, 99 minutes

Pulse Count

Adjustable, up to 9,999,999.

4-20mA

Adjustable, up to 9,999,999 (4-20mA = 0-100 Pulses/min).

External Contact

Momentary, 25 millisecond dry contact closure.

PROGRAM TYPE

Composite Terminate after up to 9,999,999 samples.

Multi-Composite

Adjustable, up to 99 cycles per bottle.

Consecutive

Adjustable, up to 9 bottles per cycle.

Daily Cycle

Adjustable, up to 9 bottles per day.

Timed Step

Adjustable, up to 99 hours, 99 minutes per step.

CLOCK

Real Time Clock

Real time operating system.

PINCH VALVE

Sample release

Adjustable, normally open / normally closed.

PURGE CYCLE

Draw and purge time

Adjustable, 1 to 99 seconds.

SUCTION CYCLE

Variable

Adjusts automatically to double the value of the purge time setting or
until liquid contacts level electrode in metering chamber.

Vacuum

System pressure range is -14 psi to +20 psi, which can be shown on
the Optional Pressure Gauge.

ALARM OUTPUTS

Independent

Container Full (Latched. Any key resets. NPN*)

Sample Fault (Latched. Any key resets. NPN*)

Cycle Abandoned (Pulsed. NPN*)

*NPN (sinking) – see Technical Appendix for details.

STATUS OUTPUTS

Independent

Sample Taken (DC relay driver, sinking)

DIRECT FUNCTION
KEYS

Manual Sample

Samples manually when pressed twice. Does not interrupt program.

Manual Purge

Purges system during second press as long as button is pressed.

Manual Bottle Advance

Moves distributor arm to next bottle.

Restart

Re-initiates program when pressed twice.

AVAILABLE
DISPLAYS

Real-Time Clock

Process Timing

Elapsed, remaining.

Process

Totals

Pulse

Counting

Internal/external.

Event

Response

With time stamp.

Multi-Level
Descriptions

Flashing

Text

AUTOMATIC
DISPLAYS

Container Full

Sample program complete.

Fault

Program not completed.

Power Interrupt –
Program Resumed

Alternating

Time

Stamp

Cycle(s)

Abandoned