Using the Windows Scanner Plugin

Software Reference for Scanners

Commands sent to the scanner will be preceded by a ">" symbol and data received
from the scanner will be preceded by a "<" symbol.

Calibration

The lamp should be calibrated when the scanner is first unpacked from the

shipping carton. Thereafter, you can calibrate whenever you change the lamp, or if
you move the scanner to a new location. The bulb may change position in transit
and the distribution of light must be adjusted to.

> Note

The

Calibrate Lamp

button should be used sparingly.

Calibrating the lamp will save data in the scanner's EEPROM memory. This

type of memory will retain data when power is not supplied, but can only be
written to about 10,000 times. Therefore, lamp calibration should NOT be a daily
procedure.

Before calibrating the unit, any film or film holders must be removed from

the scanner. If this is not done, the scanner will misinterpret the light balance and
attempt to calibrate to a setting that is outside normal limits. Next, close the dust
cover, press the

Calibrate Lamp

button and wait until the busy light goes out and

you are returned to the Scanner Tools dialog. This process will take a few minutes
on the LS-3500 and up to 30 seconds on the LS-3510AF. Do not disturb the
scanner or it may not calibrate properly.

The Scanner Tools dialog is closed by pressing the

Done

button.

Scan Size and Resolution

The LS-3510AF total scanning area is a 5000 pixel by 5000 pixel scan

covering 40 x 40 millimeters. This makes for a scan resolution of 3175 pixels per
inch or 125 pixels per millimeter. All lower resolution scans are derived from this.

Fig. 59

Scan Area of the LS-3510AF

A standard 35mm frame has a scannable area of approximately 36mm by

24mm; or 4500 pixels by 3000 pixels at highest resolution. The full scan area is