(C) Example: Programming the Monitor for Four (4) Periods

Suppose that the start time of the monitoring session is 09:00 and you would like the monitor to

measure the patient's blood pressure every 30 minutes until 22:00, and from 22:00 to 23:00 you

want the monitor OFF so that the patient may sleep, and from 23:00 to 06:00 you would like the

patient's blood pressure measured once an hour, and finally from 06:00 you want the blood pressure

measured every 15 minutes. Remember that the Time for Period One (1) also defines the End of the

monitoring session. The end time for Period Four (4) will default to the Time for Period One (1), in

this case 09:00.

PERIOD

TIME

INTERVAL

DURATION

1

09:00-22:00

30 minutes

13 hours

2

22:00-23:00

OFF

1 hours

3

23:00-06:00

1 hour

7 hours

4

06:00-09:00

15 minutes

3 hours

Total:

24 hours

The following is the recorder setup according to these instructions.

Note: Set Periods One (1), Two (2), Three (3) and Four (4). Leave the other Periods as

“-”.

PERIOD

START

INTERVAL

1

9

30 minutes

2

22

OFF

3

23

1 hour

4

6

15 minutes

5

-

-

6

-

-

Figure 25: Monitor programming example 3