51

3-2 Communication Errors

When communicating between the monitor and the PC, the following dialog box may appear if the

communication setup is not performed properly:

Figure 34: Communications Error dialog box

If this error occurs, complete the steps below:

Step 1

Click the OK button on the Error dialog box.

Step 2

Check the monitor's battery. Recharge the battery if necessary. Or replace with new
batteries.

Step 3

Check the Default Serial Port selection. Make sure it is correct.

Step 4

Check the Default Recorder Type selection. Make sure it is correct.

Step 5

Disconnect the monitor from the processor and reconnect them. (TM-2420/TM-2421)
Disconnect the cable and reconnect it. (TM-2430/UA-767PC)

Step 6

Attempt communication. You may need to go through the process a few times to re-
establish communication.

If the error still persists, complete the steps below:

Step 1

Exit the software.

Step 2

Disconnect the monitor from the processor and reconnect them. (TM-2420/TM-2421)
Disconnect the cable and reconnect it. (TM-2430/UA-767PC)

Step 3

Charge the monitor's battery. (TM-2420/TM-2421 only)

Step 5

Start the software.

Step 6

Check the Default Serial Port selection. Make sure it is correct.

Step 7

Check the Default Recorder Type selection. Make sure it is correct.

Step 8

Attempt communication again.