Factory default – BrandTech Scientific Titrette User Manual

Page 50

Advertising
background image

> 3 s

50

What to do

How to do it

Keys to press

Display readout

The continuously displayed CAL symbol indicates that the factory default setting has been changed.
If you wish to restore the factory default setting, complete the following instructions.

1.

Bring up the

CAL mode

When CAL appears in the
display, release the CLEAR
key. The input symbol blinks
and the adjustment value
for the previously accepted
adjustment appears.

Press the On/Off key and
Pause key at the same time
in order to delete the CAL
symbol.

2.

Restoring the

factory default
setting

Factory Default

CAL Mode · Easy Calibration Technique

With the instrument powered
on, press and hold the
CLEAR key for more than
3 seconds. The following
modes will be repeatedly
shown in the display after
one another:
CAL - GLP - APO - dP.