Troubleshooting – BrandTech Scientific Titrette User Manual

Page 64

Advertising
background image

64

Problem

Possible cause

Corrective action

Troubleshooting

Liquid is above the piston

Piston leaks

Perform a cleaning, replace the
piston/cylinder assembly
(see page 56).

Piston difficult to move

Piston/cylinder assembly is
contaminated or damaged by
crystalline deposits

Perform a cleaning, replace
the piston/cylinder assembly, if
necessary (see page 56).

Filling not possible

Filling valve stuck

Clean the filling valve. If the
valve ball is stuck use a 200 µl
plastic pipette tip to loosen it
(see page 60).

Filling not possible /
liquid is drawn back into the
titrating tube during filling

The discharge valve is
contaminated or the titrating
tube has been damaged

Clean the discharge valve or
exchange the titrating tube
(see page 56).

Air bubbles in the instrument

Instrument filled too quickly

Fill instrument slowly

Filling tube is loose or dam-
aged

Fasten the telescoping filling
tube firmly. If necessary, cut the
tube off approx. 1 cm from the
top or replace it.

Filling valve is loose or the
seal possibly has not been
inserted

Check whether the seal has been
inserted, and fasten the valve
securely with the mounting tool.

Filling tube does not dip into
the liquid

Fill up the bottle, or correctly
adjust the length of the telescop-
ing filling tube.

Recirculation tube is not
mounted or mounted
improperly

Attach the recirculation tube.
The opening must point outward
toward the bottle wall.

Titrating not possible

Discharge valve stuck

Clean or exchange the titrating
tube with its integrated discharge
valve (see page 56).

The volume delivered is
smaller than that indicated

The instrument has not been
completely primed

Prime the instrument again
(see page 41).

Seal might not have been
inserted or the filling valve is
loose

Check whether the seal has been
inserted, and fasten the valve
securely with the mounting tool.

Filling valve is blocked or
damaged

Clean, and if necessary replace
the filling valve (see page 60).

The instrument doesn’t
indicate any function

Internal error

Perform a restart: remove the
batteries, wait 1 minute and then
replace them (see page 61).