Limitations of use – BrandTech Scientific Transferpette S User Manual

Page 30

Advertising
background image

The instrument is intended for the pipetting of liquids within the following
limitations:
– Temperature of both the instrument and solution should be between

+15 °C to +40 °C (59 °F to 104 °F). Consult the manufacturer for use
in temperatures outside of this range.

– Vapor pressure up to 500 mbar
– Viscosity: 260 mPa s (260 cps)

Limitations of Use

The pipette is an air-displacement system for pipetting aqueous solutions
with medium density and low to medium viscosity.

Purpose

Viscous and highly adhesive liquids may impair volumetric accuracy.
Volumetric accuracy may also be impaired when pipetting liquids that
differ from ambient temperature by more than ± 1 °C /± 1.8 °F.

Operating Limitations

The user has to ensure the compatibility of the instrument with the
intended application.
This instrument cannot be used:
• for liquids incompatible with polypropylene
• for liquids incompatible with polycarbonate (inspection window)
• for liquids of a very high vapor pressure

Operating Exclusions

Limitations of Use

30

Advertising