• 5 ml and 10 ml instruments should only be used with the PE filter
installed (see page 40).
• Pipette tips are disposables items!

1.

Fitting the tip

Use the correct tip according to the volume range
or the color code. Ensure that the tip is securely
seated.

Pipetting

2.

Volume setting

a) Push the volume-change protection upward to

disengage (UNLOCK).

b) Select the desired volume by rotating the volume-

setting wheel. Avoid twisting and abrupt rotating
motions during this adjustment.

c) Push the volume-change protection down to

re-engage (LOCK). Note: The volume-change
protection tightens but does not lock volume-
setting wheel.

3.

Aspirate sample

a) Press pipetting button to the first stop.
b) Hold the pipette vertically and immerse

the tip into the liquid.

32

Volume
range

Immersion

depth in mm

Waiting

time in s

0.1 µl - 1 µl

1 - 2

1

> 1 µl - 100 µl

2 - 3

1

> 100 µl - 1000 µl

2 - 4

1

> 1000 µl

3 - 6

3