Pipetting – BrandTech Scientific Transferpette S User Manual

Page 33

Advertising
background image

English

4.

Discharge sample

a) Place the pipette tip against the wall

of the vessel. Hold the pipette at an
angle of 30-45° relative to the con-
tainer wall.

b) Press the pipetting button slowly to

the first stop and hold it down.

For serum and liquids of high viscos-
ity or low surface tension, observe
adequate waiting time to improve
accuracy.

c) The blow-out stroke empties the tip

completely: Press the pipetting button
down to the second stop.

d) While doing this, wipe the pipette tip

against the wall of the container.

e) Remove the pipette tip from the con-

tainer wall and let the pipetting button
slide back.

Pipetting

33

c) Let the pipetting button slide back

slowly. In order for the liquid to reach

its end position, leave the tip immersed

for a few seconds.