Troubleshooting – BrandTech Scientific Transferpette S User Manual

Page 46

Advertising
background image

Troubleshooting

Problem

Possible cause

Corrective action

Tip dripping
(instrument leaks)

Unsuitable tip

Only use high-quality
tips

Tip not seated tightly

Press tip on firmly

The instrument does
not aspirate or as-
pirates too little; the
discharged volume
is too low.

Seal contaminated

Clean seal

The seal or cone is
damaged

Replace seal or shaft

The piston is contami-
nated or damaged

Clean or replace piston

Aspiration is
too slow

Shaft clogged

Clean shaft

The filter in the 5 ml
and 10 ml models is
contaminated

Change the filter

Discharged volume
is too large

Pipetting button
pressed too far into
the blow-out position
before sample uptake

Operate properly.
See 'Pipetting',
page 32.

Piston is difficult
to move

The piston is contami-
nated or needs oil

Clean and oil the piston

46