Setup configuration, 1 power failure mode, 2 audible alarm enable – Next Advance SP570-M User Manual

Page 13: 3 ttl i/o operational trigger configuration

Advertising
background image

NEXT>>>ADVANCE

www.nextadvance.com

SP500 series OEM Programmable Syringe Pump

NE500 series manual

8

Revision 101101

auto-alarm message will be sent to an attached computer.

Sending any command will stop the buzzer and clear the alarm. When the problem causing the pump
motor to stall has been corrected, the Pumping Program can be resumed by any start trigger, TTL input, or
RS-232.

6. Setup Configuration

The following configuration parameters can be set or queried:

The values for "n" are "0" for off, "1" for on

Command

Description

PF n

Power Failure mode, where ‘n’ is the current setting.

AL n

Alarm mode, where ‘n’ is the current setting.

TRG aa

Operational Trigger configuration, where ‘aa’ is the current configuration.

DIN n

Directional control input setup

ROM n

Pump motor operating TTL output configuration

6.1 Power Failure Mode

Command: "PF <n>" <n> Setting: ‘0’ = Disabled, ‘1’ = Enabled.

When enabled, if the Pumping Program was operating when power to the pump was disrupted, the
Pumping Program will automatically start operating when power is reconnected to the pump.

CAUTION: The Pumping Program will start operating from the beginning of the Pumping

Program (Phase 1), regardless of what part of the Pumping Program was operating when the power was
disrupted.

When the SP500 series syringe pump is used as a component in an automated infusion/withdrawal
dispensing system, a Pumping Program can be designed to automatically synchronize the pusher block at
the start of the Pumping Program. This would be accomplished using attached sensors that send signals to
the Pumping Program.

6.2 Audible Alarm Enable

Command: "AL <n>", <n> Setting: ‘0’ = Disabled, ‘1’ = Enabled.

When alarms are enabled, the buzzer will be sounded as follows:

Condition

Buzzer Action

Pumping Program ended

Continuous beeping

Pumping Program paused for start trigger

Continuous beeping

Alarm condition, such as pump motor stalled

Steady alarm

Any command sent to the pump will stop the alarm.

6.3 TTL I/O Operational Trigger Configuration

Command: "TRG <aa>", <aa> as below in settings.

Configures how the TTL I/O ‘Operational Trigger’ (pin 2) will control the Pumping Program’s operation.
(See sec. 9.1, TTL I/O Operational Controls).

Advertising
This manual is related to the following products: