SP500 series OEM Programmable Syringe Pump

NE500 series manual

16

Revision 101101

Phase

Function

Rate

Volume

Direction

3

INC

1.0

0.1 ml

Infuse

Phase

Function

4

LOP 50

Phase

Function

5

LPS

Phase

Function

Rate

Volume

Direction

6

DEC

1.0

0.1 ml

Infuse

Phase

Function

7

L0P 99

Phase

Function

Rate

Volume

Direction

8

DEC

1.0

0.1 ml

Infuse

Phase

Function

9

LPS

Phase

Function

Rate

Volume

Direction

10

INC

1.0

0.1 ml

Infuse

Phase

Function

11

LOP 50

Phase

Function

12

JMP 02

7.4.4 Example 4: Complex Dispenses with External Synchronization

0

0

750

Syringe

Refill

A more complex dispensing example, this Program contains different pumping requirements, including
dispenses with multiple pumping rates. The first set of 3 dispenses drops down to a lower pumping rate
during the dispense. When each dispense is completed, the buzzer beeps to alert the operator, then the
pump waits for a start trigger before starting the next dispense.

The next set of 3 dispenses have a fixed time interval of 60 seconds between dispenses. After the last set
of dispenses, the syringe is refilled by the amount infused, 17.25 ml. Then the buzzer beeps, to alert the
operator to the start of the first set of dispenses. The process is then repeated.

Phase

Function

Rate

Volume

Direction

1

RAT

750.0 ml/hr

0.5 ml

Infuse

Phase

Function

Rate

Volume

Direction

2

RAT

300.0 ml/hr

1.5 ml

Infuse