SP500 series OEM Programmable Syringe Pump

NE500 series manual

28

Revision 101101

PUMPING DIRECTION

DIR [ INF | WDR | REV ]

Set/query pumping direction

INF

= Infuse

WDR

= Withdraw

REV

= Reverse pumping direction

Applicable with all Program Phase functions. Cannot be set when the Pumping Program is
operating and the “Volume to be Dispensed” is non-zero.

The pumping direction cannot be changed if an alarm condition exists.

Query response:

{ INF | WDR }

8.5.2 Pump Operational Commands

START PUMPING PROGRAM

RUN

Starts the Pumping Program operation.

If the Pumping Program was paused, then the Pumping Program resumes at the point where it was
stopped. Otherwise, the Pumping Program starts from Phase 1.

The pump cannot be started if an alarm condition exists.

STOP PUMPING PROGRAM

STP

If the Pumping Program is operating, the pump will be stopped and the Pumping Program will be
paused.

If the Pumping Program is paused, the stop command will cancel the pause and reset the Pumping
Program.

VOLUME DISPENSED

DIS

Queries volume dispense only. Set not applicable.

Response:

I <float> W <float> <volume units>

Where: “I <float>“ refers to the infusion volume dispensed, and “W <float>” refers to the
withdrawn volume.

CLEAR VOLUME DISPENSED

CLD { INF | WDR }

Sets the Infused or withdrawn volume dispensed to 0. Command is only valid while the Pumping
Program is not operating.

INF

= Infusion volume

WDR

= Withdrawn volume

Query is not applicable.