Next Advance SP570-M User Manual

Page 23

Advertising
background image

NEXT>>>ADVANCE

www.nextadvance.com

SP500 series OEM Programmable Syringe Pump

NE500 series manual

18

Revision 101101

This example demonstrates a Pumping Program whose control depends on an external sensor. Assuming a
pressure sensor that is configured to detect a high pressure point and a low pressure point, the Pumping
Program individually selects whether it will react to the high or low pressure point.

The “Program Output” pin on the TTL I/O connector (pin 5) is used to select the high or low pressure
point. When low, the low pressure point is selected (PHASE 01), and when high, the high pressure point
is selected (PHASE 05). The Program begins by infusing continuously at 10.0 ml/hr (PHASE 02), while a
background trap is set for the low pressure point (PHASE 03). To create a delay when the pressure sensor
is switched from high pressure to low pressure when the “Program Output” pin is set, a small volume is
pumped (PHASE 02, 06) before the background traps are set.

When the low pressure trap is triggered, the pump sets the high pressure trap (PHASE 07) and begins to
increment the flow rate. The flow rate is incremented in 1.0 ml/hr steps with every 0.25 ml dispensed
(PHASE 08-10). If the high pressure trap hasn’t as yet been triggered, the flow rate will max out at 25.0
ml/hr while waiting for the high pressure trap (PHASE 11). When the high pressure point is reached, the
pump immediately will drop down to 10.0 ml/hr (PHASE 02), and once again wait for the low pressure
point.

Phase

Function

1

OUT 0

Phase

Function

Rate

Volume

Direction

2

RAT

10.0 ml/hr

0.005

Infuse

Phase

Function

3

EVN 05

Phase

Function

Rate

Volume

Direction

4

RAT

10.0 ml/hr

0.0 ml (off)

Infuse

Phase

Function

5

OUT 1

Phase

Function

Rate

Volume

Direction

6

RAT

10.0 ml/hr

0.005

Infuse

Phase

Function

7

EVN 01

Phase

Function

8

LPS

Phase

Function

Rate

Volume

Direction

9

INC

1.0

0.25 ml

Infuse

Phase

Function

10

L0P 14

Phase

Function

Rate

Volume

Direction

11

RAT

25.0 ml/hr

0.0 ml (off)

Infuse