NEXT>>>ADVANCE

www.nextadvance.com

SP500 series OEM Programmable Syringe Pump

NE500 series manual

33

Revision 101101

Pin #

Definition

Type

Function

1

Vcc (5V)

Reference

Logic high reference

2

Operational Trigger

Input

Configurable start/stop operational trigger input.

Foot Switch

Falling edge:

Start or stop trigger

Level

Falling edge:

Stop trigger

Rising edge:

Start trigger

Start only

Falling edge:

Start trigger

3

Pumping Direction

Input

Changes pumping direction according to setup
[dr:rE]

[dr:dU]

Falling edge: Infuse

Withdraw

Rising edge: Withdraw

Infuse

4

Event Trigger

Input

Event input or user definable input

5

Program Output

Output

Program controlled output or user definable output

6

Program Input

Input

Program conditional input or user definable input

7

Pump Motor Operating

Output

[RUN.0] High: Pumping; Low: Not pumping

[RUN.1] High: Pumping or Pause timer
Low: Not pumping

8

Pumping Direction

Output

High: Infuse; Low: Withdraw

9

Ground (0V)

Reference

Logic low reference

9.1 TTL I/O Operational Controls

While the user is changing settings or configuration from the keypad, external control by the ‘Pumping
Direction’ and ‘Operational Trigger’ inputs will be ignored. These controls will also be ignored if an
alarm condition exists.

Operational Trigger (Pin 2): The input signal on this pin controls the operation of the Pumping
Program. Its functionality is user configurable. Use the ‘TRG' command to configure this input pin (See
Section 06.3, TTL I/O Operational Trigger Configuration).

Each option, Foot Switch, Level Control, and Start Only, define when the Operational Trigger input is
activated. When activated, the trigger emulates a ‘Start/Stop’ key.

Foot Switch:

Operates like a ‘Start/Stop’ key, whereby each falling edge (contact to ground) either
starts or stops/pauses the Pumping Program.

Level Control: Falling edge stops/pauses the Pumping Program, Rising edge starts the Pumping

Program. This configuration can be used with a contact closure timer or in an
automation setup, allowing logic level control over the operation of the pump.

Start Only:

Falling edge starts the Pumping Program. This configuration only allows the starting of
the Pumping Program. This would be useful, for example, with a laboratory animal
trained to press a lever. The animal can start the Pumping Program, but repeated presses
would have no affect until the Pumping Program permits it.

Pump Motor Operating (Pin 7): This output provides an external signal indicating when the pump
motor is operating. This pin is configured with the ROM <n> setup command. When set to 0, the output
is only at logic high when the motor is operating (pumping). When set to 1, the output is logic high when
the motor is operating or when the Pumping program is executing a pause timer. Otherwise, the output is a
logic low.