Quick start instructions – Next Advance SP570-M User Manual

Page 2

Advertising
background image

NEXT>>>ADVANCE

www.nextadvance.com

SP500 series OEM Programmable Syringe Pump

NE500 series manual

i

Revision 101101

Quick Start Instructions

Quick start instructions assume that the pump was not previous programmed with a multiple
Phase Pumping Program.

Plug in the pump.

Attach the computer cable to the "Computer" connection on the RJ-11 connector

Start the terminal emulator program on the computer

Baud Rate: 19,200
Data frame: 8N1

Note: Each command sent to the pump must have a carriage return (0x0D) at the end, but space
characters are optional and will be ignored.

Note: Wait for the response from each command before sending the next command

Setup Pumping Parameters

Set the Syringe Inside Diameter:

Where "nn.nn" is the syringe inside diameter: Send the command: DIA nn.nn

Set the Pumping Rate.

Where "nn.nn" is the pumping rate and "xx" is the rate's units: Send the command: RAT nn.nn xx

Rate units "xx" value
mL/hr

MH

µL/hr

UH

mL/mn

MM

µL/mn

UM

If the pumping rate is out of range the response message to the computer will include "?OOR"

Set the Volume to be Dispensed or Continuous Pumping

Where "nn.nn" is the volume to be dispensed, send the command: VOL nn.nn

If the volume to be dispenses is 0, then the pump will pump continuously.

Note: The units for the volume to be dispensed will be mL or µL, depending on the syringe size.
At or below 14.00 mm syringe diameter, volume units are µL
Above 14.00 mm syringe diameter, volume units are mL

Set the Pumping Direction

For infusion, send the command:

DIR INF

For Withdrawal, send the command:

DIR WDR

Load the Syringe

Press in the white drive-nut button to move the pusher block.

Insert the syringe plunger in the pusher block slot.

Insert the syringe barrel flange in the flange brackets with the syringe barrel holder on the syringe.
Tighten the flange brackets onto the syringe flange. Tighten the pusher block screw.

Operating the Pump

Start the Pump

Send the command: RUN

Stop the Pump

Send the command: STP

Query the Pump Status

Send just a carriage return to the pump. The pump will respond with a status prompt

When Pumping

The pumping rate can be changed.

With continuous pumping, the pumping direction can be changed.