Additional settings of pipettes calibration mode – RADWAG BALANCES 3Y SERIES User Manual

Page 117

Advertising
background image

- 117 -

1. Setup – access to balance’s menu

2. Print header

3. Print footer

4. Select pipette

The grey coloured workspace contains the following data:

o

Pipette

o

Tested volume

o

Average volume

o

Systematic error

o

Random error

o

Status

24.1. Additional settings of pipettes calibration mode

The additional settings enable adjusting the working mode to user’s needs and

requirements. To access the setting follow below procedure:

Procedure:

1. Press grey coloured workspace,

2. The display indicates menu of: Settings, Keys, Information, Printout

and Profile,

3. Press <Settings> key,

4. The display indicates functions related to calibration of pipettes.

Pipettes calibration is related to the following functions:

Number of measurements; this option enables determining number of

measurements for tested volume of a pipette (valid for each tested

volume in case of a pipette with adjustable volume),

Request serial number: YES/NO. For option <YES> before starting

calibration procedure, the software displays a window and on-screen

keyboard for inserting serial number of a pipette,

Comply with ISO 8655: YES/NO. For option <YES>, for tested

volumes, during pipettes calibration process the software automatically

accepts error values as specified in standard ISO 8655 (if other error

values have been accepted for a pipette, then on selecting option

Advertising