Additional settings of parts counting mode – RADWAG BALANCES 3Y SERIES User Manual

Page 72

Advertising
background image

- 72 -

16.1. Additional settings of parts counting mode

The additional settings enable adjusting the working mode to user’s needs and

requirements. To access the setting follow below procedure:

Procedure:

1. Press grey coloured workspace,

2. The display indicates menu of: Settings, Keys, Information, Printout and

Profile

3. Press <Settings> key

4. The display indicates functions related to weighing and parts counting

modes.

Parts counting mode features the following optional functions:

ACAI, Automatic Accuracy Correction:

o

YES, mass of a single part is updated during counting process

o

NO, mass of a single part is not updated

Means of operation of ACAI function:

1. Number of parts (on adding) on balance’s weighing pan has to be

greater than has been previously.

2. Number of parts (on adding) on balance’s weighing pan must be less

than twice the amount of which was visible on the display before adding

parts.

3. Current quantity of parts must be within the ± 0,3 tolerance of the total

value.

4. Measurement result has to be stabilized.

Minimal reference mass: 1unit, 2units, 5units, 10units, it is the

minimal mass value of a single part. Unless the condition is met, counting

process cannot start.

Result control:

o

YES, print and save only those measurements which are included

within the Lo and Hi thresholds.

o

NO, all measurements are printed and saved.

Other functions of menu <Settings>:

Tare mode

Automatic footer printout