RADWAG BALANCES 3Y SERIES User Manual

Page 44

Advertising
background image

- 44 -

ONLINE – selecting this option downloads current air density value

from a connected THB sensor, if such is connected to the balance,

or from internal sensor installed in the balance.

If a balance features both types of sensors (THB module and

internal sensor), then data from the THB module supersedes the

other, and parameters Setup/Ambient Conditions/Ambient

Conditions Module require setting to SAVE or SAVE & ALERTS

mode.

If a balance features only one type of the sensors for temperature,

humidity and pressure recording, then correct ONLINE operation

requires setting of balance parameters:

Internal sensors only - Setup/Ambient Conditions/Ambient

Conditions Module set to NONE,

External THB module only - Setup/Ambient

Conditions/Ambient Conditions Module set to SAVE or

SAVE & ALERTS mode.

Additionally, for correct balance cooperation with the THB module

set port’s transmission parameters to compatible with those of the

THB module. the THB module transmission parameters are detailed

on its data plate.

On returning to the weighing mode, the display indicates an additional symbol
<

!

>, as indicated on below drawing.From now on the value of indicated mass

is corrected including the air buoyancy compensation factor and sample

density.

The measurement result can only be compensated by a correct value if an

operator inserts to balance’s memory the actual value on the air density and

correct density of weighed sample.

Caution:

If the <AIR BUOYANCY COMPENSATION> mode is set to <ONLINE> value,

then balance display does not indicate pictogram <!> which means that

parameter Setup/Ambient Conditions/Ambient Conditions Module is

set to SAVE or SAVE & ALERTS mode, and there is no physical connection

between a balance and a THB module, or the parameters for communicating

the module with the balance are set incorrectly. In such case plug a THB

module to balance’s ports COM 1 or COM 2 and set appropriate cooperation

parameters, such as specified on module’s data plate.