29.11. Reports on pipette calibration processes

Reports contain data on completed pipette calibration processes. In case of

multichannel pipettes, the reports are generated for each channel separately.

Each report can be previewed, searched by date, exported or printed.

Procedure

Enter submenu < Databases>, and press <

Reports from pipette

calibration processes >

Press field of a desired report record, unless it is visible on the top of the

list, wind the list forward using navigating keys

Report’s name consists of date and time of its creation,

e.g.: 2012.03.12 11:12:15

Caution:

It is possible to use an option for searching a report.

List of parameters in a report on pipette calibration process:

1. Pipette [name from the database of pipettes]

2. Serial no. [serial no. inserted during calibration process]

3. Channel no.

4. Start date

5. End date

6. User [name of a user carrying out calibration]

7. Client [client name]

8. Number of measurements

[no. of measurements for each tested

volume]

9. Operation with ISO 8655

[information if errors comply with the

standard]

10. Status [information whether the errors are below permissible

value accepted for the tested volume]

11. Temperature [average temperature value during calibration

process]

12. Humidity [average humidity value during calibration process]

13. Pressure [average pressure value during calibration process]

14. Water temperature [average water temperature during

calibration process]

15. Z coefficient [value if the Z coefficient for given temperature]

29.12. Series

Database of series contains data saved for a series and measurements of

samples.

Procedure

Enter submenu < Databases>, and press <

Series> field

Press <

Add > key to add a new series to the database