Autozero function – RADWAG WPT/P 60 Postal Scales for Packages User Manual

Page 23

Advertising
background image

12.3. Autozero function

The autozero function has been implemented in order to assure precise
indications. This function controls and corrects „0” indication. While the
function is active it compares the results continuously with constant
frequency. If two sequentional results differ less than the declared
value of autozero range, so the scale will be automatically zeroed
and the pictograms

and

will be displayed.

When AUTOZERO is disabled zero is not corrected automatically.
However, in particular cases, this function can disrupt the measurement
process e.g. slow pouring of liquid or powder on the weighing pan. In this
case, it is advisable to disable the autozero function.

Procedure:

• Enter the submenu <P1.rEAd> and then:

Fnnd

no

- filter disabled

Fnnd

YES - filter enabled


Return to weighing:

See - 10.2.2.

- 23 -